Enhanced adaptive matched filter for automated identification and measurement of electrocardiographic alternans
Abstract
Electrocardiographic alternans, consisting of P-wave alternans (PWA), QRS-complex alternans (QRSA) and Twave alternans (TWA), is an index of cardiac risk. However, only automated TWA measurement methods have been proposed so far. Here, we presented the enhanced adaptive matched filter (EAMF) method and tested its reliability in both simulated and experimental conditions. Our methodological novelty consists in the introduction of a signal enhancement procedure according to which all sections of the electrocardiogram (ECG) but the wave of interest are set to baseline, and in the extraction of the alternans area (AAr) in addition to the standard alternans amplitude (AAm). Simulated data consisted of 27 simulated ECGs representing all combinations of PWA, QRSA and TWA of low (10 mu V) and high (100 mu V) amplitude. Experimental data consisted of exercise 12-lead ECGs from 266 heart failure patients with an implanted cardioverter defibrillator for primary prevention. EAMF was able to accurately identify and measure all kinds of simulated alternans (absolute maximum error equal to 2%). Moreover, different alternans kinds were simultaneously present in the experimental data and EAMF was able to identify and measure all of them (AAr: 545 mu V x ms, 762 mu V x ms and 1382 mu V x ms; AAm: 5 mu V, 9 mu V and 7 mu V; for PWA, QRSA and TWA, respectively) and to discriminate TWA as the prevalent one (with the highest AAr). EAMF accurately identifies and measures all kinds of electrocardiographic alternans. EAMF may support determination of incremental clinical utility of PWA and QRSA with respect to TWA only.