Screening for refractive errors in children: The plusoptiX S08 and the Retinomax K-plus2 performed by a lay screener compared to cycloplegic retinoscopy

Abstract

PURPOSE To evaluate the performance of the autorefractor Retinomax K-plus2 and the photo-screener plusoptiX S08 in measuring refractive errors by comparing them with cycloplegic retinoscopy (CR) and to assess limitations associated with their use. METHODS Cross-sectional study to compare data from CR, performed by an orthoptist, to data from Retinomax K-plus2 and plusoptiX S08 performed by a lay screener. Sensitivity and specificity for the detection of significant refractive errors were determined according to American Academy of Pediatric Ophthalmology and Strabismus criteria. RESULTS Two hundred children were included, with a mean age of 5.2 +/- 2.6 years (3 months to 11 years). Compared to CR, the plusoptiX S08 showed a mean difference of -1.13 +/- 1.25 D (95% limits of agreement [LOA], -3.59 to +1.32) for spherical equivalent (SE) and -0.23 +/- 0.53 D (LOA, -1.28 to +0.81) for the cylinder. Mean difference for the Retinomax K-plus2 before cycloplegia was -0.08 +/- 0.58 D (LOA, -1.23 to +1.06) for SE and 0.03 +/- 0.38 D (LOA, -0.72 to +0.78) for the cylinder; after cycloplegia -2.11 +/- 1.64 D (LOA, -5.33 to +1.10) for SE and -0.06 +/- 0.47 D (LOA, -0.98 to +0.86) for the cylinder. Sensitivity for detecting hyperopia >3.5 D with the plusoptiX S08 was 33.3%, the Retinomax before cycloplegia 31.0% and after cycloplegia 84.6% and high for detecting myopia, astigmatism, and anisometropia. CONCLUSIONS Retinomax K-plus2 and plusoptiX S08 have high sensitivity for the detection of myopia, astigmatism, and anisometropia compared to cycloplegic retinoscopy; however, when used without cycloplegia, hyperopia is underestimated. (J AAPOS 2010;14:478-483)

vision screener preschool-children adult patients down-syndrome autorefraction amblyopia accuracy young photorefraction agreement